Sep 27, 2026News & Insights

How to Plan a Medical Pendant for an Endoscopy Room: A PLQJ Checklist

Use this checklist to match PLQJ arm reach, rotation, platforms, load and service connections to an endoscopy room plan before requesting a configuration.

plqj-single-arm-endoscopic-pendant
A medical pendant for an endoscopy room should be planned from the equipment stack, mounting position and required connections—not selected from the room name alone. The PLQJ Single Arm Endoscopic Pendant provides a 700–1100 mm customizable arm, separate or simultaneous arm and terminal-box rotation, three adjustable platforms and a ≥220 kg net load. Each of these specifications should be checked against the actual room plan before a configuration is requested.
This checklist turns the published PLQJ specifications into practical questions for hospital procurement teams, planners and medical equipment distributors.

Start with the endoscopy room and equipment map

Begin with a dimensioned drawing that shows the procedure table, endoscopy equipment, planned pendant mounting point and other major fixed items. Mark the intended working position of the equipment stack and the area that the pendant must reach.
The PLQJ Single Arm Endoscopic Pendant has a horizontal arm radius of 700–1100 mm, which is customizable. The published arm thickness is at least 10 mm. Because the arm range is a radius from the mounting point, the mounting coordinate and the desired equipment position need to be considered together.
Do not assume that the longest available arm is automatically the correct choice. Instead, use the room drawing to confirm whether the proposed radius places the terminal box and platforms where the equipment is intended to stand. The final layout should also be reviewed with the project's qualified parties responsible for the ceiling structure and mounting solution.

Translate PLQJ specifications into planning questions

The most useful way to read a product specification is to connect every number to a project question.
Verified PLQJ specification
Question to confirm before ordering
700–1100 mm customizable horizontal arm radius
What distance is required between the proposed mounting point and the planned equipment position?
0–340° horizontal rotation
Which positions must the pendant reach within the room plan?
Arm and terminal box can rotate separately or simultaneously
Does the proposed orientation require the box to turn independently from the arm?
Net load ≥220 kg
What is the combined weight of all equipment, accessories and items to be supported?
Three adjustable 560 × 550 mm platforms
Do the equipment dimensions fit the platform area, and is a three-level arrangement suitable?
Eight 10A power sockets
How many powered devices are listed, and what plug and electrical requirements apply at the destination?
One network interface and two grounding terminals
Does the project connection schedule require these quantities or a customized arrangement?
One drawer and one infusion pole with four hooks
Are these accessories required in the planned workflow and position?
The table is a planning aid, not an installation drawing. Confirm the final configuration against the equipment list, room drawing and project utility schedule.

Check movement and working position

PLQJ's horizontal arm and terminal box can rotate separately or simultaneously through 0–340°. This published movement is useful only when it is matched to the proposed mounting point and the positions where the equipment will be used.
Use the room plan to review:
  • The primary working position of the endoscopy equipment
  • Any secondary position that the pendant must reach
  • The orientation required for platform access and connections
  • The relationship between the pendant, procedure table and other fixed equipment
  • The position in which the pendant will be kept when the equipment is not in use
This exercise helps the supplier understand the requested coverage. It also makes it easier to determine whether the proposed arm radius and rotation arrangement match the project rather than relying on a general description such as “standard endoscopy room.”

Verify platform size, quantity and total supported load

PLQJ includes three platforms with adjustable height. Each platform is listed as 560 × 550 mm and includes silicone-gel corner protection with an anti-collision design.
Prepare the width, depth, height and weight of every item intended for the platforms. The ≥220 kg net-load specification is important, but total load is not the only check. Equipment dimensions, platform quantity and the vertical arrangement also matter.
A useful equipment schedule can include:
  • Equipment name or model
  • Width, depth and height
  • Individual weight
  • Intended platform level
  • Required power connection
  • Required network connection, if any
  • Cables or accessories that affect positioning
Do not count only the largest device. Add the planned equipment and accessories together, then discuss the complete list with the supplier.

Confirm power, network, grounding and gas requirements

The published PLQJ accessory list includes eight 10A power sockets, one network interface, two grounding terminals and one drawer. Its working power supply is listed as AC110V or 220V at 50Hz.
For gas interfaces, German, European, British and other standards can be customized. The published product information also states that interface colors and shapes differ for anti-wrong connection, the number of plug and unplug cycles exceeds 20,000, and the interfaces use secondary sealing with on, off and unplugged states.
Before requesting a quotation, provide a connection schedule that lists:
  • Destination electrical supply and plug requirements
  • Quantity and location of power sockets
  • Required network interfaces
  • Required grounding terminals
  • Gas types and quantities
  • Applicable national or project gas-interface standard
  • Preferred interface positions on the terminal box
“Customizable” does not mean that the supplier can infer the correct standard or quantity. The required connections should be stated explicitly in the inquiry.

Review the listed accessories

PLQJ includes one stainless-steel adjustable infusion pole stand with four infusion-bottle hooks. It also lists one drawer. Confirm whether these items fit the intended room arrangement and whether the proposed side and height are appropriate for the project.
Accessory requirements should be connected to the equipment layout. For example, the selected platform levels, connection locations, drawer position and infusion-pole position should be reviewed together rather than as separate items.

When another medical pendant layout may fit better

PLQJ is specifically presented as a single-arm endoscopic pendant, but a different verified model may fit a different room requirement.
  • The PLDT Single Arm Pendant has a 700–1100 mm customizable arm, ≥220 kg net load and two adjustable 600 × 450 mm platforms. It is a useful comparison when a general two-platform arrangement is being considered.
  • The PLDQ Pendant Bridge uses a 2500–3000 mm beam, subject to on-site measurement, with separate dry and wet sections and a ≥450 kg net load. It addresses a wider service-zone requirement rather than one single-arm workstation.
  • The PLEMT Electrical Anesthesia Pendant includes an electric anesthesia-machine lifting bracket with a 300 mm lifting stroke. It is the relevant model to review when the main requirement is anesthesia-machine positioning.
All four products are available on the Medical Pendant category page. Compare the room requirement first, then the model details.

What to include in an endoscopy pendant inquiry

A complete inquiry should contain enough information for the configuration to be discussed against the actual project. Send:
  1. Room purpose and a dimensioned layout
  1. Proposed ceiling mounting coordinate
  1. Required working position and arm coverage
  1. Equipment list with dimensions and weights
  1. Preferred platform arrangement
  1. Power, network and grounding schedule
  1. Gas types, quantities and required interface standard
  1. Drawer and infusion-support requirements
  1. Destination electrical supply
  1. Any site-measurement or project drawing requirements
If some information is not yet final, label it as preliminary. This makes it clear which points still need confirmation.

Frequently asked questions

Is PLQJ specifically intended for endoscopy rooms?

Yes. The published product name is PLQJ Single Arm Endoscopic Pendant. The final configuration should still be matched to the specific equipment list, room drawing and connection schedule.

How many equipment platforms does PLQJ have?

PLQJ has three adjustable platforms. Each platform is listed as 560 × 550 mm.

Can the arm and terminal box rotate independently?

Yes. The published specification states that the horizontal arm and terminal box can rotate separately or simultaneously through 0–340°.

Is the ≥220 kg net load enough to select the pendant?

No. The load rating is one check. Equipment dimensions, platform arrangement, arm radius, mounting position and connection requirements also need to be reviewed.

Which gas-interface standard should be ordered?

PLQJ can be configured for German, European, British and other customizable standards. State the standard required by the destination project in the inquiry.
To review a PLQJ configuration for a specific endoscopy room, contact YIKUN Medical and send the room plan, equipment schedule and required connections.