NT-6600
Detail

Throughput

60 test/hour

Assay Items

5 parts, 28 parameters, 2 histograms, 2 scattergrams

Test Mode

CBC+5DIFF, CBC

Measuring Principle

Laser Scatter Technology

Sample Modes

Autoloader: 50 tubes, auto mixing, random direction tube barcode scanning technology

Close tube: STAT priority, support both whole blood and capillary blood samples

Testincj Items

Regular Parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW.SD, RDW-CV, PLT, PDW,

MPV. PCT. P-LCR. NEU#t NEU%. LYM#m LYM%. MONff. EOS#. EOS%. BAS#. BAS%


Research Parameters: ALY#, ALY%, IG#, IG%

Linearity Range

Item

Linear range

Carry over

cv

WBC

1.0-99.9-10VL

<0.5%

<2.0%

RBC

0.3-7.0,10,7/L

<0.5%

51.5%

HGB

20-2400/L

<0.5%

<1.5%

PLT

20-999'10*/L

<1.0%

S4.0%

Storage

100,000 results storage including histogram and scattergram

US Interface

LIS Interface with HL7 protocol

QC and Calibration

Multiple QC rules, including L-J, X-B etc. Auto and manual calibration funtion

Incubation

Integral titanium incubation system

Multi-warning

Reagent inventory check; Abnormal original data warning; Multi warning flags

Power Supply

AC110/220V±10%, 60/50HZ

Package Size(W*D*H)

810*760*850mm

Gross Weight

104kg